Criticare Non-Invasive Blood Pressure Monitors
VitalCare 506DX
new model

VitalCare {Model 506DX} Specifications
Pulse Rate Display and Alarms
Source: SpO2 or NIBP
Resolution: 1 BPM
Range: 20-300 BPM
Accuracy: Greater of ±1% or ± 1 BPM
Unspecified for <40 or>290 BPM
Non-Invasive Blood Pressure
Technique: Automatic oscillometric upon inflation
Pressure Range: Adult mode: 30-300 mmHg
Neonate mode: 30-150 mmHg
Maximum Inflation Pressure: 300 mmHg
Accuracy: Greater of ±2% or ±2mmHg over full range
Measurement Cycle: <40 sec avg. (0-300 mmHg: Standard Adult Cuff) Automatic Measurement Cycles: OFF, 1, 2, 3, 5, 10, 15, 30, 45 or 60 minutes, 2 or 4 hrs; STAT function initiates 5 minutes of consecutive NIBP measurements Oxygen Saturation (SpO2)
Averaging Time: Selectable, 3, 6, 9, 12, 15, 18, or 21 seconds
Search Time: Selectable, 10, 20, 30, or 40 seconds
Sensor Type: Reusable or disposable
Range: 1-99%
Resolution: 1%
Accuracy: ± 2% (70-99%); ±3% (40-69%); unspecified below 40%
Method: Dual-wavelength LED
Temperature (506DXNT only)
Range: Normal mode: 32.2°C to 42.2°C
(90.0°F to 108.0°F)
Monitor mode: 26.7°C to 42.2°C
(80.0°F to 108.0°F)
Resolution: ±0.1°C (±0.2°F)
Accuracy: ±0.1°C (±0.2°F) When tested in a calibrated water bath (in monitor mode)
Battery
Type: Rechargeable lead-acid (sealed)
Rating: 6V, 4 A-H
Life: 6 to 8 hours
Recharge Time: 12 hours to 80% capacity
16 hours to 100% capacity
Power Requirements
• Do not connect the monitor directly to any external power source!
• In order to ensure compliance with applicable IEC standards,
use only the CSI Cat. Nos. 905 or 908 charger.
Charger/transformer,
Charger Input Rating: Cat No. 905: 225 to 240V, 50/60Hz
Cat No. 908: 110 to 120V, 50/60Hz
Cat No. 902-1: 100 to 130V, 50/60Hz
Operating Environment
Temperature: 10° to 40° C (50° to 104° F)
Humidity: 15 to 90% (noncondensing)
Transport and Storage Environment
Temperature: -5° to 50° C (23° to 122° F)
Humidity: 15 to 90% (non condensing)
Atmospheric Pressure: 619 to 775 mmHg (825 to 1033 hP)
Display
Type: LCD , LED
Physical
Size: 6"H x 5" W x 5" D (15.24cm. H x 12.7cm. W x 12.7 cm. D)
Weight: 4.4 lbs. (2.2 kg)
Type of Protection: Class I Equipment
Degree of Protection: Type CF Equipment (SpO2, temperature)
Type CF Equipment, Defibrillator-Proof (NIBP only)
Protection Against Liquids: IPX 1 rating, Drip-Proof Equipment
Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air, oxygen or nitrous oxide.
Mode of operation: Continuous operation
Note: Specifications subject to change without notice.
Performance of the monitor may be adversely affected by exposure to high levels
of electrostatic discharge.
|