Criticare Non-Invasive Blood Pressure Monitors VitalCare 506DX

Criticare Non-Invasive Blood Pressure Monitors
VitalCare 506DX

new model

Criticare Non-Invasive Blood Pressure Monitors VitalCare 506DX

VitalCare {Model 506DX} Specifications


Pulse Rate Display and Alarms
	Source:	SpO2 or NIBP
	Resolution:	1 BPM
	Range:	20-300 BPM
	Accuracy:	Greater of ±1% or ± 1 BPM 
			Unspecified for <40 or>290 BPM
	

Non-Invasive Blood Pressure
	Technique:	Automatic oscillometric upon inflation
	Pressure Range:	Adult mode:		30-300 mmHg
			Neonate mode:	30-150 mmHg
	Maximum Inflation Pressure:	300 mmHg
	Accuracy:	Greater of ±2% or ±2mmHg over full range
	Measurement Cycle:	<40 sec avg. (0-300 mmHg: Standard Adult Cuff) Automatic Measurement Cycles: OFF, 1, 2, 3, 5, 10, 15, 30, 45 or 60 minutes, 2 or 4 hrs; STAT function initiates 5 minutes of consecutive NIBP measurements Oxygen Saturation (SpO2)
	Averaging Time:	Selectable, 3, 6, 9, 12, 15, 18, or 21 seconds
	Search Time:	Selectable, 10, 20, 30, or 40 seconds
	Sensor Type:	Reusable or disposable
	Range:		1-99%
	Resolution:	1%
	Accuracy:	± 2%  (70-99%); ±3% (40-69%); unspecified below 40%
	Method:		Dual-wavelength LED

Temperature (506DXNT only)
	Range:		Normal mode:	32.2°C to 42.2°C
					(90.0°F to 108.0°F)	
			Monitor mode:	26.7°C to 42.2°C
					(80.0°F to 108.0°F)
	Resolution:	±0.1°C (±0.2°F)
	Accuracy:	±0.1°C (±0.2°F)	When tested in a calibrated water bath (in monitor mode)

Battery
	Type:		Rechargeable lead-acid (sealed)
	Rating:		6V, 4 A-H
	Life:		6 to 8 hours
	Recharge Time:	12 hours to 80% capacity
			16 hours to 100% capacity

Power Requirements 

•	Do not connect the monitor directly to any external power source! 
•	In order to ensure compliance with applicable IEC standards,
	use only the CSI Cat. Nos. 905 or 908 charger.

	Charger/transformer, 
	Charger Input Rating:	Cat No. 905: 225 to 240V, 50/60Hz
				Cat No. 908: 110 to 120V, 50/60Hz	
				Cat No. 902-1: 100 to 130V, 50/60Hz			

Operating Environment
	Temperature:	10° to 40° C (50° to 104° F)
	Humidity:	15 to 90% (noncondensing)

Transport and Storage Environment
	Temperature: 	-5° to 50° C (23° to 122° F)
	Humidity:	15 to 90% (non condensing)
	Atmospheric Pressure: 	619 to 775 mmHg (825 to 1033 hP)


Display 
	Type:	LCD , LED

Physical
	Size:	6"H x 5" W x 5" D   (15.24cm. H x 12.7cm. W x 12.7 cm. D)
	Weight:	4.4 lbs. (2.2 kg)
	Type of Protection:	Class I Equipment
	Degree of Protection:	Type CF Equipment (SpO2, temperature)
					Type CF Equipment, Defibrillator-Proof  (NIBP only)
	Protection Against Liquids:	IPX 1 rating, Drip-Proof Equipment

	Equipment not suitable for use in the presence of a flammable
	anesthetic mixture with air, oxygen or nitrous oxide.
		
	Mode of operation:	Continuous operation


Note:  Specifications subject to change without notice.

	Performance of the monitor may be adversely affected by exposure to high levels
	of electrostatic discharge.

 


Criticare Non-Invasive Blood Pressure Monitors VitalCare 506DX